HOME PRODUCTS SERVICES COMPANY
INFO
SPECIALS CONTACT
US

Topcon SL-D Series Digital Slit Lamps
Ready for Digital Imaging The Topcon SL-D Series is the first true digital slit lamp series that combines high optical performance with versatile digital capability, The SL-D Series incorporates excellent optics that assure optimal visualization of all structures of the eye. The instruments also incorporate unique filters to further enhance their clinical efficiency. The concept of the "true" digital slit lamp is based on the extensive range of optional digital capture devices that can be mounted on these outstanding clinical instruments. Paramount among these is the dedicated DC-1 Digital Camera that provides easy-to-capture, high quality digital images. To further advance their "digital-ready" attractiveness, an auto left/right eye detection function is incorporated into the slit lamps. These premier diagnostic instruments offer superior clinical performance with unmatched digital capability.

The Digital Camera DC-1 (Optional Accessory) is built exclusively for the SL-D Series to provide superb digital documentation. A variety of image capture modes can be taken including still or video images. The DC-1 integrates the beam splitter, PC card storage and the high-resolution digital camera into one unobtrusive and compact system. The images captured by the DC-1 can be quickly downloaded to any PC imaging software package from a compact flash card or be directly transmitted via a USB cable. For more efficient patient database management, all of the SL-D Digital Slit Lamps can be connected to Topcon's IMAGEnet Imaging Systems.

•  Easy-capture withjoystick button on slit lanip
•  Still images and digital video recording
•  2X and 4X zoom function
•  Download JPEG images via compact flash card or USB cable
•  Compact and user-friendly design
•  Low-flash requirement for patient comfort
•  Multi language menu selection


Click here to download more info

Features
Photo Attachment (SL-D7 and SL-D8Z Models (Optional Accessory)
The Topcon Photographic Attachment is specially designed for the SL-D7 and SL-D8Z to allow various DSC/CCD cameras to be used with the unique FD-21 Flash Device. The entire illumination used for image capturing is produced by a high-intensity flash tube with very short exposure time to provide superb images and additional details. Superior clarity, color-balance, resolution and depth-of-field, assist in accurate examinations. The Photographic Attachment accepts and supports an extensive range of DSC/CCD cameras, such as:

•  Topcon Digital Camera DC-1
•  F-mount Digital Camera Back
•  Sony CCD Camera DXC-990P/390P
•  Other 1/2" and 1/3" Mount CCD Cameras

IMAGEnet Systems (option)
The Topcon SL-D Series is capable of inputting images and data into the Topcon IMAGEnet family of software products for versatile diagnostic processing and interpretation. The digital images generated can be captured for use in research, education, and clinical applications. IMAGEnet2000 is a full-function, state-of-the-art digital imaging system for faster, more efficient acquisition, storage, retrieval and analysis of all types of ophthalmic images.

Fluorescein Observation of the Anterior Segment (SL-D7 and SL-D8Z Models Factory Option)
When requested at the time of order. a special combination of optional filters can be incorporate to provide the eye care specialist with excellent fluorescein images of the anterior segment. These efficient filters aid the practitioner in detecting abnormalities of the corneal and conjuntival epithelium while permitting the use of lower levels of light intensity for greater patient comfort.

Amber Filter
The SL-D Series incorporates an amber filter, which can be easily inserted or removed and improves the contrast and color depth of the images.

SL-D2 Three Magnifcation Digital Slit Lamp
The SL-D2 incorporates a parallel drum magnification changer that allows for quick and easy selection among three convenient magnifications of 10X, 16X and 25x. The SL-D2 offers the most cost-effective digital imaging solution with optimal clarity, color resolution and depth-of-field.

SL-D4 Five Magnification Digital Slit Lamp
Five Magnification Digital Slit Lamp The SL-D4 incorporates a parallel drum magnification changer that allows for quick selection among five convenient magnifications of 6X, 10X, 16X, 25X and 40X. The 40X magnification is especially useful for Cortical Endothelium observation. Encompassing a variety of ophthalmic applications, the demanding eye care specialist can produce excellent images of both the anterior and posterior structures of the eye.

SL-D7 Five Magnification Digital Slit Lamp
The SL-D7 incorporates a parallel drum magnification changer that allows for quick selection among five convenient magnifications of 6X, 10X, 16X, 25X and 40X. When the optional FD-21 Flash Device is incorporated, the SL-D7 converts to a'true' digital photographic slit lamp providing exceptional images of the cornea, endothelium and fundus to enhance accurate documentation.

SL-D8Z Zoom Magnification Digital Slit Lamp
The SL-D8Z offers a zoom magnification changer, which provides continuous observation through the magnification range of 6X to 32X.The highly efficient optics allow for extraordinary visualization of minute anterior segment conditions. When the optional FD-21 Flash Device is incorporated, the SL-D8Z becomes the premier zoom photographic slit lamp for unparalleled digital documentation.

Specifications
Microscope SL-D2 SL-D4 SL-D7 SL-D8Z
Type Galilean magnification changer with converging binocular tubes
Magnification Selection 3 steps by
drum rotation
5 steps by
drum rotation
5 steps by
drum rotation
Variable via continuous zoom
(click 10X, 16X, 25X)
Eyepiece 12.5X
Magnification ratio (Field of View) 10X (22.5mm)
16X (14.1mm)
25X (8.8mm)
6X (35.1mm
10X (22.5mm)
16X (14.1mm)
25X (8.8mm)
40X (5.6mm)
6X (35.1mm
10X (22.5mm)
16X (14.1mm)
25X (8.8mm)
40X (5.6mm)
Continuous from 6X to 32X
(35.2mm to 7mm)
Slit Width Continuous from 14mm to 0mm (at 14mm, slit becomes a circle)
Slit Length Continuous from 14mm to 1mm (at 14mm, slit becomes a circle)
Slit Angle 0 to 180 degree with horizontal scanning capability
Filters Blue, Red-Free, Amber
UV Cut (Normal use) IR Cut (Normal Use)
Blue, Red-Free, Amber
UV Cut (Normal use) IR Cut (Normal Use)
ND (13% Transmission), Heat Insulation
Exciter / Barrier (Factory Option)
Illumination Lamp 6V, 20W halogen lamp 12V, 30W halogen lamp
Longitudinal Movement (Base) 90mm
Lateral Movement (Base) 100mm
Fube Base Movement 15mm
Vertical Movement (Base) 30mm
Vertial Movement (Chin rest) 80mm
Fixation Target Red LED Red LED or White Bulb
Power Input 100-120V, 220-240V AC
Frequency 50 / 60Hz
Power Consumption 160VA
Dimensions (w/table) 550mm x 399mm x 558mm 550mm x 399mm x 760-790mm
Dimensions (w/unit table) 440mm x 379mm x 558mm 440mm x 399mm x 760-790mm
Dimensions w/o table and chinrest 329mm x 306mm x 415-445mm 329mm x 306mm x 652-682mm
Weight (w/table) 17kg 19kg 19.5kg
Weight (w/unit table) 16kg 18kg 18.5kg
Weight (w/o table and chinrest) 9.5kg 11.5kg 12kg